PSY Firmware Release Notes: - R1-7-8: Fixes measurement scheduling issues. Added verification of C1, C2 temperature correction factors. Added checks to address issue with positive water potential readings. Increased stability in accessing the configuration in flash memory. - R1-7-1: • Corrected instances where a ',' is written instead of the ';' as a delimiter, ignoring regional settings - R1-7-0: Stability Improvements added for measurement schedules to decrease errors - R1-6-9: Adds improvements to: a) Micro SD Card Data Management (Driver and File System Stability Updates) b) Micro SD Power Control Management Improvements c) Data File header and corruption fixes such as missing lines in data or complete blocks missing - R1-6-4: Fixes an issue which prevented file uploads from PSY to the web via the MCC hub - R1-6-2: a) Improvements to stability of data storage on SD card. b) Fixed bug where SD card marked as “In Use” on start-up if SD card is not present due to a flag not being cleared. - R1-6-1: Fixes an issue when downloading a .csv data file using ICT Combined Instrument Software. - R1-6-0: a) Improvements to SD card power management. b) Improvements to microcontroller data management. c) Internal battery voltage and external power supply is logged by default and cannot be changed by the user. d) Update to raw data file format to fix issue with incorrectly aligned time after cooling.